Recall Z-2322-2012 ยท IZL System, X-Ray, Mobile

Siemens Medical Solutions USA, Inc

Siemens Mobilett Mira. Mobile X-Ray System.

DateSep 6, 2012 (initiated Aug 21, 2012)
StatusTerminated
ReasonDuring regular product monitoring, Siemens has become aware of a potential issue with the Mobilett Mira system. During normal clinical operation the switch on the motor drive handle of the unit may become loose and slide horizontally to the right and become trapped under the cover. This may cause the system to continuously move. In response, Siemens has issued Update Instruction XP038/12/S to r
Root causeDevice Design
ActionSiemens initiated their correction of the MOBILETT Mira on August 21, 2012 by sending an "UPDATE INSTRUCTIONS XP038/12/S to all affected customers. The Instruction identifies the product, problem, and work steps for customers to follow. A XP038/12/S Completion Protocol letter was included for customers to complete and file with the instruction once the update is completed.
Quantity20
DistributionNationwide Distribution- including the states of GA, ID, IL, IA, KS, NC, OH, SD, TX, and WI.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K111912MOBILETT MIRA
Siemens Medical Solutions USA, Inc., Sep 19, 2011