ClearanceBrief
Recall Z-2432-2012 ยท GDW Staple, Implantable

Covidien LP

Covidien Duet TRS 45 3.5 mm Universal Straight Single Use Loading Unit. Product Code: DUET4535. Single Use Loading Unit (SULU) tissue reinforcement stapler. The Endo GIA Universal Staplers with Duet TRS reloads have applications in abdominal, gynecologic and pediatric surgery for resection, transection and creation of anastomoses. They may be used for transection and resection of liver substance

DateSep 21, 2012 (initiated Aug 21, 2012)
StatusTerminated
ReasonDuet TRS may have the potential to injure adjacent anatomical structures which may result in life threatening post-operative complications
Root causeDevice Design
ActionCovidien issued an "URGENT MEDICAL DEVICE RECALL" letter dated August 21, 2012 to all affected customers. The letter identified the product, problem, and actions to be taken by the customers. The firm sent a revised customer recall notification letter via FedEx overnight delivery to all customers who purchased Duet TRS" devices from August 2009- August 2012. This letter went to all customers whether they had been sent the August 21st letter or not. Users were advised that Covidien was recalling all lots of the Duet TRS Universal Straight and Articulating Single Use Loading Units (SULU). On
Quantity433,528 units
DistributionWorldwide Distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K080898AUTO SUTURE ENDO GIA STAPLERS WITH ENDO GIA SINGLE USE LOADING UNITS WITH STAPLE LINE REINFORCEMENT
Tyco Healthcare Group LP, May 5, 2008