ClearanceBrief
Recall Z-2428-2012 ยท HAW Neurological Stereotaxic Instrument

Integra LifeSciences Corporation

Integra CRW Precision Arc (CRWPRECISE), CRW Arc System (CRWASL), HRAIM Head Ring Assembly with Intubation Mounts The CRW stereotactic system is used in conjunction with the HRAIM head ring is a target-centered system that can be configured to be CT-only or CTIMR compatible. The CRW stereotactic system is a multi-purpose system used for localizing intercranial targets for precisely directing inst

DateSep 20, 2012 (initiated Aug 9, 2012)
StatusTerminated
ReasonT-handle screw is used on a complementary product to the CRW, the Integra HRAIM, which is the CT-compatible intubation Head Ring Assembly. When the T-handle on the HRAIM intubation hoop is tightened, the T-handle could stop in a vertical position that prevents the CRWPRECISE and certain models of the CRW-ASL from seating properly on the head ring although it appears to be fully seated
Root causeDevice Design
ActionThe firm, Integra, sent an "URGENT: VOLUNTARY MEDICAL DEVICE RECALL NOTIFICATION" dated August 10, 2012 to US & Canada consignees/customers via traceable courier service and Non-US & Non-Canada consignees/customers - through written notification delivered either by traceable courier, email, facsimile, or through a directed visit by an Integra Sales Representative. The notification describes the product, problem and actions to be taken. The Integra sales representative will replace the screws according to the attached rework instructions. If for some reason it is not possible to replace the sc
Quantity175 units
DistributionWorldwide distribution: USA (Nationwide) and countries of: Argentina, Australia, Belgium, Canada, China, Czec, Denmark, France, Hong Kong, Hungary, Italy, Japan, Malaysia, Philippines, Singapore, Spain, South Africa, and UK.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K944463CRW-1 SYSTEM
Radionics Software Applications, Inc., Oct 26, 1995