Recall Z-0160-2013 ยท DTZ Oxygenator, Cardiopulmonary Bypass

Maquet Cardiovascular Us Sales, Llc

QUADROX iD HMOD 2030 Diffusion Membrane Oxygenator; QUADROX-i ADULT HMO 70000 / HMO 71000; QUADROX-iD Pediatric HMOD 30000 Product Usage: The QUADROX oxygenators are intended for the use in extracorporeal circulation during cardiopulmonary bypass in cardiac surgery. Within the specified flow rate range, the device oxygenates the blood, eliminates carbon dioxide and regulates blood temperature. T

DateOct 26, 2012 (initiated Jun 12, 2012)
StatusTerminated
ReasonIt has come to firm's attention that, in some rare cases, the blood outlet connector on some models of MAQUET Oxygenators may detach from the oxygenator. This event has occurred in about 0.01% (1:10,000) of units shipped.
Root causeNonconforming Material/Component
ActionMaquet sent a Field Safety Notices and Confirmation Forms dated June 13, 2012 to all affected customers. A Follow-up communication dated July 20, 2012 was mailed to all affected US customers. The Field Safety Notice identifies the problem, affected product, risk factors and actions to be taken by the customer. Customers were instructed to complete the Confirmation form and return in the self-addressed prepaid envelop. For questions call 201-995-8968. Maquet issued an updated Field Safety Notice/Field Safety Notice Confirmation Form dated 2/14/13 via Fed Ex on 2/18/2013. Maquet sent out a
Quantity23,203 units
DistributionUS Nationwide Distribution
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K082117QUADROX-I ADULT WITH AND WITHOUT INTEGRATED ARTERIAL FILTER WITH SOFTLINE COATING, MODEL# HMO 70000, HMO 71000
Maquet Cardiopulmonary, AG, Feb 18, 2009
K090511QUADROX-I ADULT MICROPOROUS MEMBRANE OXYGENATOR WITH BIOLINE COATING, MODEL BEHMO70000,BEQ-HMO70000
Maquet Cardiopulmonary, AG, Jun 12, 2009
K090689QUADROX-I SMALL ADULT WITH AND WITHOUT INTERGRATED ARTERIAL FILTER WITH SOFTLINE COATING, MODELS HMO 50000, HMO 51000
Maquet Cardiopulmonary, AG, Oct 29, 2009
K093522QUADROX-I SMALL ADULT MICROPOROUS MEMBRANE OXYGENATOR; W/ BIOLINE COATING AND W/ INTEGRATED ARTERIAL FILTER W/ BIOLINE C
Maquet Cardiopulmonary, AG, Dec 7, 2009
K103191QUADROX-I PEDIATRIC MICROPOROUS MEMBRANE OXYGENATOR WITH INTEGRATED ARTERIAL FILTER WITH SOFTLINE / BIOLINE COATING
Maquet Cardiopulmonary, AG, Mar 8, 2011
K112360QUADROX-I MICROPORPOROUS MEMBRANE OXGENATOR SERIES AND QUDROX-ID DIFFUSION MEMBRANE OXYGENATORS SERIES
Maquet Cardiopulmonary, AG, Sep 15, 2011