Recall Z-2256-2012 ยท IZL System, X-Ray, Mobile

GE Healthcare, LLC

Optima XR200amx. Mobile general purpose radiographic imaging of the human head and body.

DateSep 13, 2012 (initiated Jul 10, 2012)
StatusTerminated
ReasonSoftware issue: a violation of 21 CFR 1020.31(a). After an image is taken, when the user selects and clears an additionally displayed screen the system inhibit is also cleared, allowing the user to take an exposure without the technique factors being displayed.
Root causeRadiation Control for Health and Safety Act
ActionThe firm will issue Important Electronic Product Radiation Warning letters that identify the affected product along with a description of the defect and related hazards. Users are instructed to ensure that the acquisition mode is always selected prior to attempting an exposure to confirm the techniques selected. The firm will remedy the issue or bring the device into compliance free of charge. A GE Healthcare Service Representative will update the software on the system to address the issue. Field Modification Instructions (FMI) 10865(2) describes the rework plan the GEHC Engineers will
Quantity254 total in the US
DistributionWorldwide Distribution -- USA (nationwide) Distribution