Recall Z-0438-2013 ยท HTO Reamer

Extremity Medical LLC

IO FiX X-Post Reamer, Extremity Medical, www.extremitymedical.com; Customer Service: 888.499.0079 The X-Post Reamer is an accessory to the IO FiX Screw and Washer System that is used for reaming bone most common to orthopedic surgical procedures.

DateNov 26, 2012 (initiated Apr 25, 2011)
StatusTerminated
ReasonProduct Event Reports revealed a broken reamer experienced intra-operatively. Pre-drilling was not performed prior to reaming.
Root causeEmployee error
ActionExtremity Medical sent a notification by e-mail that describes the "Slight modification to the Surgical Technique" on April 25, 2011. For questions regarading this recall call 973-588-8980.
Quantity156 units
DistributionWorldwide Distribution - USA including FL, MA, NC, CA, MD, VA, IN, AL, NJ, AZ, OK, MT, NY, OH, IA, MI, TX, WI, GA, IL, TN and NE. Internationally to Switzerland.