Recall Z-2405-2012 ยท IYE Accelerator, Linear, Medical

Elekta, Inc.

Precise Digital Accelerator Delivery of radiation to defined target volumes

DateSep 19, 2012 (initiated Aug 6, 2012)
StatusTerminated
ReasonThere have been a small number of instances in hospitals where the joint between the drive gear and the gantry base drive wheel has become loose, leading to fatigue failure of some of the securing bolts.
Root causeNonconforming Material/Component
ActionElekta sent an Important Field Safety Notice dated August, 2012 to inform users that inspection and possibly bolt replacement needs to be carried out. Users should return the signed acknowledgement to Elketa, and an inspection should be carried out under FCO20002103053. For questions customers should call +44(0)1293 654200. For questions regarding this recall call 770-300-9725.
Quantity110
DistributionNationwide Distribution including AL, AZ, CA, CO, CT, FL, GA, IL, IA, MD, MA, MI, MN, NE, NV, NJ, NM, NY, ND, OH, OK, OR, PA, VA, SC, TN TX, WA, WV, WI and Puerto
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K051932ELEKTA SYNERGY; ELEKTA SYNERGY S AND XVIR3.5
Elekta, Inc., Aug 12, 2005