Recall Z-2340-2012 ยท IYE Accelerator, Linear, Medical

Elekta, Inc.

Multileaf Collimator To be used for single or multiple fractions, delivered as static and/or dynamic beams of radiation, in all areas of the body where such treatment is indicated.

DateSep 10, 2012 (initiated Jul 27, 2012)
StatusTerminated
ReasonInformation released with the product Agility did not originally contain information on compatibility with the add-on Dynamic Multileaf Collimator Apex.
Root causeError in labeling
ActionELEKTA sent an Important Field Safety Notice, FCO Ref: 20001405000 dated July 7, 2012, to current Agility users and a separate notice FCO Ref: 20001407026 for current Apex users who may be in the process of considering an upgrade to Agility. Both notices made it clear that Apex has not been integrated or classified as compatible for use with Agility. Users are advised not to use Apex with Agility. For questions regarding this recall call 770-300-9725.
Quantity2
DistributionNationwide Distribution including DC & MD
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K121328AGILITY
Elekta , Ltd., Jun 12, 2012