ClearanceBrief
Recall Z-2390-2012 ยท GKP Instrument, Coagulation, Automated

Instrumentation Laboratory Co.

Instrumentation Laboratory ACL TOP 700 LAS PN 0000280030 ACL TOP is a bench top, fully automated, random access analyzer designed specifically for in vitro diagnostic clinical use in the hemostasis laboratory for coagulation and/or fibrinolysis testing in the assessment of thrombosis and/or hemostasis. The systems provide results for both direct hemostasis measurements and calculated parameters

DateSep 14, 2012 (initiated Aug 3, 2012)
StatusTerminated
ReasonLow Abnormal Fibrinogen Sample Reports as Normal and High Fibrinogen Sample Reports as Low
Root causeSoftware design
ActionThe firm, Instrumentation Laboratory, sent a "Mandatory Test Parameter Upgrade" letter dated June 7, 2012 with upgrade kits P-14.8 via USPS on August 3 thru August 10 to US and Canadian customers and notified International Affiliates via e-mail on August 2, 2012. The customers were ask to complete and return the Mandatory Test Parameter Upgrade Tracking Form via fax: 781-861-4207 or email: ra-usa@ilww.com. If you have any questions, call 781-861-4204.
Quantity46 units
DistributionWorldwide distribution: USA (nationwide) and countries of: ARGENTINA, AUSTRALIA, AUSTRIA, BELARUS, BELGIUM, BRAZIL, BRUNEI DARUSSALAM, BULGARIA, CHILE, CHINA, DENMARK, FINLAND, FRANCE, GERMANY, HONG KONG, HUNGARY, INDIA, IRELAND, ISRAEL, ITALY, JAPAN, KOREA, REPUBLIC OF LUXEMBOURG, MACAU, MALAYSIA,
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K091980ACL TOP 700 LAS
Instrumentation Laboratory CO, Sep 22, 2009