Recall Z-2463-2012 ยท LLZ System, Image Processing, Radiological

Imaging Sciences International, LLC

DEXIS Imaging Suite - catalog number 1.008.7932, version numbers 10.0.0 through 10.0.3 Radiological Image Processing System

DateSep 26, 2012 (initiated Jul 30, 2012)
StatusTerminated
ReasonA condition can occur during renumbering of the patient database using the DEXCopy feature. When renumbering is being performed using the DEXCopy feature and multiple patients with an identical name are present, the software will assign all subsequent records to the first patient record, resulting in mixed patient records. In order for this problem to occur several conditions need to be met: 1)
Root causeSoftware design
ActionDEXIS sent a Medical Device Corrective Action letter dated July 30, 2012, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. An Acknowledgement Form was included with the notification that is to be returned by the consignee to the recalling firm. Customers were instructed to call DEXIS Technical Support at 888-883-3947 as soon as possible to have Technical Support review and potentially correct any issues that may have resulted from its use. To request a free-of-charge upgrade kit, customers should email a request to ord
Quantity366
DistributionWorldwide Distribution - USA including Washington, DC, AL, AZ, AR, CA, CO, CT, DE, FL, GA, IL, IN, IA, KS, KY, LA, MD, MA, MI, MN, MO, NE, NV, NH, NJ, NM, NY, NC, ND, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI and the country of Canada.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K090461IMAGE-ARENA 4.0 AND IA APPLICATIONS 2D CARDIAC PERFORMANCE ANALYSIS 1.0
Tomtec Imaging Systems GmbH, May 22, 2009