Recall Z-2265-2012 ยท MQB Solid State X-Ray Imager (Flat Panel/Digital Imager)

Canon Inc.

Canon Solid State X-ray Imager (Flat Panel/Digital Imager), Digital Radiography, Models CXDI-55C, CXDI-55G and CXDI-70C. These devices are intended to replace radiographic film/screen systems in all general purpose diagnostic procedures.

DateAug 24, 2012 (initiated Nov 1, 2010)
StatusTerminated
ReasonScrews holding the 55REF circuit board are loosening and roaming inside of the sealed sensor.
Root causeNonconforming Material/Component
ActionCanon Inc. notified sales companies including Canon USA on 9/15/2010 with Service Bulletin 10-017, and on 8/25/2011 with Service Bulletin 11-011, and on 1/20/2011 with Service Bulletin 11-001. Sales representatives were to visit customers and check and replace the screws.
Quantity468 units
DistributionNationwide Distribution (USA)
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K091435DIGITAL RADIOGRAPHY MODEL CXDI-55G
Canon, Inc., May 29, 2009
K091436DIGITAL RADIOGRAPHY, MODEL CXDI-55C
Canon, Inc., May 29, 2009
K102012DIGITAL RADIOGRAPHY
Canon Inc. -Medical Equipment Group, Aug 17, 2010