Recall Z-2221-2012 ยท JAA System, X-Ray, Fluoroscopic, Image-Intensified

GE OEC Medical Systems, Inc

GEHC OEC 8800 Mobile Fluoroscopic X-ray System, GE Healthcare, Surgery, Salt Lake City, UT. Mobile fluoroscopic examination of human anatomy.

DateAug 17, 2012 (initiated Oct 19, 2007)
StatusTerminated
ReasonX-ray units did not meet manufacturer's specification for linearity.
Root causeNonconforming Material/Component
ActionGE updated their user manual and changed their internal specifications to match those of the CFR In addition, a software code correction was to be made to forward productions in order to correct the reported defect. *** This recall issued was previously classified under an incorrect recall number. It is being reclassified with a new number. ***
Quantity175 units