Recall Z-2253-2012 ยท MEB Pump, Infusion, Elastomeric

I-Flow Corporation

ON-Q pump with ONDEMAND bolus button, ON-Q C-bloc with ONDEMAND; ON-Q C-bloc with Select a Flow and ONDEMAND Model/Reference #: 7000180, 7000181, 7000182, 7000367, 7000397, CB003, CB005, CB006, P400X1-7+5/60, P400X2-14+5/30, P400X5+5/60, and PMB01. Part Numbers: 101347100, 101347300, 101347400, 101360200, 101368400, 101368600, 101368700, 102987000, 102987200, 102987300, 103161800, 103534300, 500

DateAug 28, 2012 (initiated May 8, 2012)
StatusTerminated
ReasonIt was determined that in a small quantity of ON-Q pumps with ONDEMAND, the bolus button may not lock in the down position when depressed and/or the orange bolus refill indicator may stay in the lowermost position. As a result, the patient may receive continuous infusion at a rate greater than expected.
Root causeDevice Design
ActionAn Important Voluntary Recall Notice letter dated 5/8/12 was sent to all their customers who purchased the ON-Q pump with ONDEMAND bolus button. The letter provided customers with the problem indicated and the actions to be taken. Customers were instructed to complete and sign the attached Recall Response Form and fax to (920)969-4565 or scan and email to IflowProductInquiry@kcc.com. Once I-Flow receives the completed FAX BACK form, an I-Flow Customer Service Representative will immediately contact the customers to provide instructions on product return and credit. Customers with questions
Quantity80,646 units total (65,926 units in US)
DistributionWorldwide Distribution -- USA, Canada, Australia, the EU (Germany), United Arab Emirates, and Cyprus.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K023318I-FLOW ELASTOMERIC PUMP WITH BOLUS
I-Flow Corp., Oct 18, 2002