Recall Z-2262-2012 ยท OXO Image-Intensified Fluoroscopic X-Ray System, Mobile

GE OEC Medical Systems, Inc

GEHC OEC Injection Cable for MedRad Provis V Contrast Injector. In certain fluoroscope X-ray modes, the cable provides a signal to the auto-injector to trigger a contrast injection. Part numbers include 00-902260-01 and 00-883654-01.

DateAug 23, 2012 (initiated Oct 10, 2008)
StatusTerminated
ReasonGE initiated a correction to the instructions/use due to the potential for erroneous auto injection of contrast dye.
Root causeDevice Design
ActionAn Urgent Safety Advisory Notice was sent to consignees on October 10, 2008. The letter identified the affected product and notified customers of the situation, along with instructing them to discontinue use until proper validation and instruction update for cable use was completed. GEHC OEC was to have contacted customers when the solution became available and resolved it at no cost to the customer.
Quantity149 units
DistributionNationwide distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K021049OEC 9800 PLUS DIGITAL MOBILE IMAGING SYSTEM
Ge Oec Medical Systems, Apr 17, 2002
K022069OEC FLUOROTRAK 9800 PLUS
Ge Oec Medical Systems, Sep 17, 2002
K024012OEC 9800 E/CV+ DIGITAL MOBILE IMAGING SYSTEM
Ge Oec Medical Systems, Mar 20, 2003
K041932OEC OLYMPUS FLUOROSCOPIC IMAGING SYSTEM
Ge Oec Medical Systems, Inc., Aug 18, 2004