Recall Z-2224-2012 ยท IQO Device, Prosthesis Alignment

Biomet U.K., Ltd.

Oxford Gap Gauge SML 3/4mm. The instrument is used for an intermediate check of the flexion and extension gaps when preparing the femur to accept the femoral component.

DateAug 17, 2012 (initiated Jul 27, 2012)
StatusTerminated
ReasonA drafting error during an unrelated design change has led to 4 misleading dimensions on the drawing. The CAD model remained the same thickness but the problem dimensions were accidentally struck from an incorrect point result in an increase of 1.0mm in the value of each dimension
Root causeDevice Design
ActionBiomet sent a "FIELD SAFETY CORRECTIVE ACTION" letter dated July 24, 2012 to all affected customers. The letter identifies the product, problem, and actions to be taken by the customers. A Response Form was attached for customers to complete and return. Contact the firm for questions regarding this notice.
Quantity119 units
DistributionWorldwide Distribution-USA-including the states of IN, SD, and TX and the countries of Austria, China, Japan, Netherlands, Sweden, Turkey, and UK.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.