Recall Z-2224-2012 ยท IQO Device, Prosthesis Alignment
Biomet U.K., Ltd.
Oxford Gap Gauge SML 3/4mm. The instrument is used for an intermediate check of the flexion and extension gaps when preparing the femur to accept the femoral component.
| Date | Aug 17, 2012 (initiated Jul 27, 2012) |
| Status | Terminated |
| Reason | A drafting error during an unrelated design change has led to 4 misleading dimensions on the drawing. The CAD model remained the same thickness but the problem dimensions were accidentally struck from an incorrect point result in an increase of 1.0mm in the value of each dimension |
| Root cause | Device Design |
| Action | Biomet sent a "FIELD SAFETY CORRECTIVE ACTION" letter dated July 24, 2012 to all affected customers. The letter identifies the product, problem, and actions to be taken by the customers. A Response Form was attached for customers to complete and return. Contact the firm for questions regarding this notice. |
| Quantity | 119 units |
| Distribution | Worldwide Distribution-USA-including the states of IN, SD, and TX and the countries of Austria, China, Japan, Netherlands, Sweden, Turkey, and UK. |