ClearanceBrief
Recall Z-2163-2012 ยท CHL Electrode Measurement, Blood-Gases (Pco2, Po2) And Blood Ph

Radiometer America Inc

The ABL software version 6.12 for the ABL800 and ABL700 analyzer. The analyzers are intended for the in Vitro testing of: -Samples of whole blood for the parameters pH, pO2, pCO2, cK+, cNa+, cCa2+, cCl-, cGlu, cLac, cCrea, ctBil, and co-oximetry parameters (ctHb, sO2, and the hemoglobin fractions FO2Hb, FCOHb, FMetHb, FHHb and FHbF) -Samples of expired air for the parameters pO2 and pCO2 -Pleura

DateAug 8, 2012 (initiated Jul 5, 2012)
StatusTerminated
ReasonA customer has reported an error on an ABL8xx configured to a Hospital Information System using the data transfer protocol HL7 version 2.5. If the instrument transmits results to a HIS system and the connection is broken the results are queued for later transmission. If the ABL and HIS system are configured with HL7 2.5, any results in the queue will be mixed up when the instrument transmits t
Root causeSoftware design
ActionThe firm, Radiometer, sent two "Priority Level: Urgent" letters one dated July 9, 2012 and the other dated July 24, 2012 to its affected customers via FedEx with tracking. The letter identifies the product, problem, and actions to be taken by the customers. The customers were instructed to complete and return the attached Recall Response Fax Form via fax to 800-736-0601. The letter informed the customers that a local service representative will be contacting them shortly to install a software update that will prevent this from occuring (in the event that you utilize HL7 version 2.5 or intend
Quantity6854 ABL 7xx units
DistributionWorldwide distribution: USA (nationwide) including the states of AK, AL, AR, AZ, BM, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, and WY, and the countries of Canada
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K002290ABL-700 SERIES, MODEL ABL 7XX (VARIOUS MODELS)
Radiometer Medical A/S, Sep 13, 2000
K041874ABL800 FLEX
Radiometer Medical Aps, Oct 8, 2004
K100777ABL837 FLEX ANALYZER
Radiometer Medical Aps, Dec 10, 2010
K110416ABL835 FLEX ANALYZER WITH PLEURAL PH
Radiometer Medical Aps, Mar 24, 2011
K111897ABL90
Radiometer Medical Aps, Jul 19, 2011
K992859ABL700 SERIES ANALYZER WITH AUTOCHECK MODULE
Radiometer America, Inc., Jan 10, 2000