ClearanceBrief
Recall Z-0284-2013 ยท DTQ Console, Heart-Lung Machine, Cardiopulmonary Bypass

Terumo Cardiovascular Systems Corporation

Perfusion System 8000, base 4 pump, 100V safety monitor included The Sarns" Modular Perfusion System 8000 is indicated for use for up to 6 hours on the extracorporeal circulation of blood for arterial perfusion, regional perfusion, and cardiopulmonary bypass procedures, when used by a qualified medical professional who is experienced in the operation of this or similar equipment.

DateNov 13, 2012 (initiated Sep 25, 2012)
StatusTerminated
ReasonTerumo Cardiovascular Systems (Terumo CVS) has received multiple reports since 2007 of various malfunctions of the arterial and cardioplegia monitors for the Sarns Modular Perfusion System 8000. The malfunctions can be broadly categorized as follows monitor power malfunctions, inability to measure pressure and inaccurate and/or false pressure alarms. If the malfunction of the arterial or cardio
Root causeDevice Design
ActionThe firm, Terumo CVS, sent an "URGENT MEDICAL DEVICE RECALL" notice/Safety Advisory Alert dated September 24, 2012 and an addendum to the Operator's Manual to alert all affected Customers. The notice described the product, problem and actions to be taken. The customers were instructed to review this medical device recall notice; assure that all users are aware to this notice; place the addendum in the Operator's Manual just before page 4.1 and complete and return the attached Customer Response Form via Fax to: 1-734-741-6149 or email to:www.termuo-cvs.com. If you have any questions or concern
Quantity26
DistributionWorldwide distribution: USA (nationwide) including states of: AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, NH, NJ, NM, NY, OH, OK, OR, PA, RI, SC, TN, TX, VA, WA, WI, WV, and WY; and countries of: ARGENTINA, AUSTRALIA, BANGLADESH, BRAZ
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K915183SARNS 8000 MODULAR PERFUSION SYSTEM
3M Health Care, Sarns, Jul 8, 1992