ClearanceBrief
Recall Z-2258-2012 ยท JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use

Beckman Coulter Inc.

UniCel DxI 600 Access Immunoassay Systems, Part Numbers: Dxl 600: A30260; Dxl 600 with spot B: A71460. The DxI Immunoassay System is an in vitro diagnostic device used for the quantitative, semi-quantitative or qualitative determination of various analyte concentrations found in human body fluids.

DateAug 23, 2012 (initiated Oct 13, 2011)
StatusTerminated
ReasonThe recall was initiated because Beckman Coulter has received seven reports of a software anomaly for the UniCel Dxl Immunoassay Systems. If undetected, this anomaly may cause erroneous test results.
Root causeSoftware design
ActionThe firm, Beckman Coulter, sent an "URGENT PRODUCT CORRECTION" letter dated October 13, 2011 with attached PCA Response Form to all customers who purchased the UniCel Dxl Immunoassay Systems and DxC Clinical Systems which includes all software versions. The letter describes the product, problem and actions to be taken. The customers were instructed verify the system temperatures immediately following each initialization; do not begin processing samples until this verification is complete; refer to the appendix included with the letter for detailed instructions about verifying system temperat
Quantity2442 units total (784 units in US)
DistributionWorldwide distribution: USA (nationwide) and countries of: Algeria, Angola, Argentina, Australia, Austria, Bangladesh, Belgium, Botswana, Brazil, Bulgaria, Canada, China, Colombia, Croatia, Czech Republic, Finland, France, French Polynesia, Germany, Greece, Hong Kong, Hungary, India, Ireland, Israel
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K023764UNICEL DXI 800 ACCESS IMMUNOASSAY SYSTEM, MODEL 973100
Beckman Coulter, Inc., Jan 28, 2003
K042291UNICEL DXC 600 AND 800 SYNCHRON SYSTEMS
Beckman Coulter, Inc., Nov 12, 2004