Recall Z-0058-2013 ยท FRN Pump, Infusion

Belmont Instrument Corporation

Belmont Rapid Infuser, a Fluid Management System, Model FMS2000.

DateOct 12, 2012 (initiated Sep 25, 2012)
StatusTerminated
ReasonLeak from a saline bag caused saline to drip along the back of the Rapid infuser close to the Power Entry Module causing an electrical flash at the power entry area.
Root causeDevice Design
ActionBelmont Instrument initiated recall on 9/25/12. A "Field Correction" notice was provided that described the product issue and recommended actions. US-based sales representatives visited all facilities to provide the modified power cord/shield. For units sold outside the United States, the modified power cords were sent to distributors who would ensure the new power cords were provided to all facilities within their accounts.
Quantity2,913 units
DistributionWorldwide Distribution including Nationwide (USA) and the countries of Australia, Austria, Belgium, Brazil, Canada, China, Denmark, England, Estonia, France, Germany, Hong Kong, India, Israel, Italy, Jordan, Mongolia, Netherlands, New Zealand, Norway, Pakistan, Peru, Puerto Rico, S. Korea, Saudi Ara
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K032674BELMONT FLUID MANAGEMENT SYSTEM, MODEL FMS2000
Belmont Instrument Corp., Sep 16, 2003
K091855THE BELMONT FLUID MANAGEMENT SYSTEM/ THE BELMONT RAPID INFUSER, MODEL FMS2000
Belmont Instrument Corp., Jul 1, 2009
K972284FLUID MANAGEMENT SYSTEM
Belmont Instrument Corp., Nov 13, 1997