Recall Z-2313-2012 ยท NBC Test, Natriuretic Peptide

Siemens Healthcare Diagnostics, Inc

Siemens ADVIA Centaur Multi-Diluent 1 - Used for automated dilutions for the ADVIA Centaur BNP and TSH3 Ultra assays when on-board the ADVIA Centaur or ADVIA Centaur XP systems. Catalog Numbers: 2 Pack REF 07907174 (110312) 6 Pack REF 07293184 (110313). Used for automated dilutions for the ADVIA Centaur BNP and TSH3 Ultra assays when on-board the ADVIA Centaur or ADVIA Centaur XP systems.

DateSep 6, 2012 (initiated Jul 25, 2012)
StatusTerminated
ReasonDecrease in onboard dilution recovery when using Multi-Diluent 1 that has been stored on board the ADVIA Centaur and ADVIA Centaur XP systems for for BNP and TSH3 Ultra assays
Root causeNonconforming Material/Component
ActionSiemens sent an "URGENT DEVICE RECALL NOTICE" dated July 2012 to all affected customers. The letter identifies the product, problem, and actions to be taken by the customers. A Field Correction Effectiveness Check Response Form was included for customers to complete and return via fax to (302) 631-7597. Contact your local Sales Representative for questions regarding this notice.
Quantity2 Packs = 15879 and 6 Packs = 7414
DistributionWorldwide Distribution-USA (nationwide) and the countries of Canada, Argentina, Australia, Brazil, Chile, China, Croatia, Denmark, Ecuduar, Egypt,Guatemala, Hong Kong,India,Ireland, Israel,India, Japan,Korea, Mexico, Malyasia, Peru, Pakistan, Saudi Arabia,Singapore, Thailand,Uruguay, Viet Nam, and S
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K031038ADVIA CENTAUR B-TYPE NATRIURETIC PEPTIDE (BNP) ASSAY
Bayer Healthcare, LLC, Jun 23, 2003
K083844ADVIA CENTAUR TSH3-ULTRA METHOD
Siemens Healthcare Diagnostics, May 8, 2009