Recall Z-2240-2012 ยท JAA System, X-Ray, Fluoroscopic, Image-Intensified

GE OEC Medical Systems, Inc

892.1650 Image-intensified fluoroscopic x-ray system; the device is designed to provide fluoroscopic and spot film imaging of the patient during diagnostic surgical and interventional procedures. The systems include features specifically designed for use in diagnostic and interventional cardiac imaging procedures, and are also intended for cholangiography, endoscopic, urologic, orthopedic, neurol

DateAug 22, 2012 (initiated Nov 6, 2009)
StatusTerminated
ReasonGE OEC is recalling certain 9800 Image Intensified Flouroscopic x-ray systems due to customer complaint analysis showing that a replacement Hitachi 160 GB Hard Drive, when installed on certain models is susceptible of causing unanticipated system shut downs, no boots, data loss or data mix, or unexpected hard drive corruption.
Root causeDevice Design
ActionGE OEC sent a Urgent Recall Notice dated November 6, 2009, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. The notification advised them that GE Healthcare had identified a potential safety issue associated with a "Significant increase on Data Loss, Data Mix, and System file corruption of the Hard Drives" on the 9800 Systems. The field correction consisted of installing a "Smart Power Switch" between the power control PCB and the system power switch on 9800 Systems with a Celeron single board computer and software ve
Quantity3,595 units
DistributionWorldwide Distribution -- USA (nationwide)
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K021049OEC 9800 PLUS DIGITAL MOBILE IMAGING SYSTEM
Ge Oec Medical Systems, Apr 17, 2002
K022069OEC FLUOROTRAK 9800 PLUS
Ge Oec Medical Systems, Sep 17, 2002
K024012OEC 9800 E/CV+ DIGITAL MOBILE IMAGING SYSTEM
Ge Oec Medical Systems, Mar 20, 2003