Recall Z-2275-2012 ยท LLZ System, Image Processing, Radiological

Siemens Medical Solutions USA, Inc

syngo.plaza, Model number 10592457. Intended use: radiological image processing system.

DateAug 27, 2012 (initiated Jul 18, 2012)
StatusTerminated
ReasonThere is a potential malfunction in which a miscalculation occurs for images with pixel depth values greater that 12 bit in certain functions.
Root causeSoftware design
ActionThe firm initiated their recall/correction of this device on July 18, 2012 by sending a "Customer Safety Advisory Notice" letter, dated July 12, 2012, to their consignees. The letter identified the affected product along with the malfunction. It also discussed the steps the user can take to avoid the issue. Users were asked to not use the mentioned functions for images from those modalities until the issue was resolved and the system was updated with the software version containing the solution. Users were to contact their local Siemens Service in order to plan for the updated software ver
Quantity14
DistributionNationwide Distribution -- CA, FL, GA, MA, MN, MO, NJ, NY, OH, WA, and WI.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K101666SYNGO. PLAZA VA20A
Siemens AG Medical Solutions, Aug 10, 2010