Recall Z-2264-2012 ยท LLZ System, Image Processing, Radiological

Siemens Medical Solutions USA, Inc

syngo Imaging XS. Radiological image processing system.

DateAug 24, 2012 (initiated Jul 18, 2012)
StatusTerminated
ReasonThere is a potential malfunction when using syngo Imaging XS, version VA70A or higher. In some situations it can happen that a miscalculation is done for the grey scale values in the functions "Regions of Interest", "Pixel Lens", "Edge Enhancement" and "Histograms".
Root causeSoftware design
ActionSiemens sent a "CUSTOMER SAFETY ADVISORY NOTICE" dated July 12, 2012 to all affected customers. The letter identifies the product, problem, and actions to be taken by the customers. For questions concerning this notice contact your local Siemens Service Rep..
Quantity52
DistributionNationwide Distribution-CA, CO, CT, FL, GA, IL, IN, IA, KS, MD, MN, MO, NE, NV, NJ, NY, NC, OH, PA, TN, TX, and WI.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K082430SYNGO IMAGING XS
Siemens AG Medical Solutions, Nov 14, 2008