Recall Z-2155-2012 ยท GAD Retractor

Stryker Spine

LUXOR RETRACTOR BASE; Catalog Number: 48250030; Manufactured by: STRYKER Spine SAS; Z.I MARTICOT-33610-CESTAS-FRANCE; Distributed in the USA by: Stryker Spine. 2 Pearl Ct. Allendale, NJ 07401. Provides access to the thoracic and lumbar spine from a posterior approach via a small incision.

DateAug 7, 2012 (initiated Apr 23, 2012)
StatusTerminated
ReasonStryker received three reports involving blades of the Luxor Retractor becoming disassembled from the base causing the retractor to collapse insitu.
Root causeNonconforming Material/Component
ActionStryker sent Urgent Product Recall letters/return response forms dated April 23, 2012 via FedEx to Stryker branches/agencies and hospitals where product might have been distributed. The letter identified the affected product, discussed the issue, potential hazards, risk mitigations, and the return process. Customers are to examine their inventory and reconcile any of the affected product with the information provided. In addition, the Customer Response Form should be completed and faxed back to Regulatory Compliance at 201-760-8370 within 5 days of receipt of the letter. The affected produ
Quantity30 devices
DistributionWorldwide Distribution -- US, Switzerland, Netherlands, Brazil, Hong Kong, Germany, and Taiwan.