Recall Z-2228-2012 ยท LIP Enzyme Linked Immunoabsorption Assay, Treponema Pallidum

Siemens Healthcare Diagnostics, Inc

Siemens ADVIA Centaur Syphilis (SYPH) Assay , 200 ReadyPack Reference Number 10492493 Product Usage: The ADVIA Centaur Syphilis (SYPH) assay is an in vitro diagnostic immunoassay for the qualitative determination of antibodies to Treponema pallidum in human serum or plasma (EDTA, lithium or sodium heparinized, citrate) using the ADVIA Centaur and ADVIA Centaur XP systems as an aid in the diagn

DateAug 20, 2012 (initiated Jul 17, 2012)
StatusTerminated
ReasonSyphilis Assay Interference with HBsAg Assay
Root causeComponent design/selection
ActionSiemens Healthcare sent Urgent Device Recall Notice letters dated July 2012 via e-mail to all affected customers inside and outside the US. The letters identifed the affected product, reason for voluntary recall and risk to health. The letters provided actions to be taken by the customers and if there are any remaining inventory of the affected product, follow the guidelines outlined. Customers were instructed to complete and return the Field Correction Effectiveness Check form. For questions contact your local Technical Support Provider or Distributor.
Quantity2050 Kits
DistributionWorldwide Distribution - US Nationwide and the countries of: Foreign: Brazil, China, Denmark, Hong Kong, Malaysia, and New Zealand Viet Nam
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K112343TREPONEMA PALLIDUM TREPONEMAL TEST REAGENTS; QUALITY CONTROL MATERIALS
Siemens Healthcare Diagnostics, Inc., Jan 20, 2012