Recall Z-2175-2012 ยท LEH Radioimmunoassay, Vancomycin

Siemens Healthcare Diagnostics, Inc.

Dimension Vista(R) Vancomycin Flex(R) Reagent Cartridge (VANC, Catalog # K4086). For the quantitative measurement of vancomycin in human serum and plasma on the Dimension Vista system.

DateAug 9, 2012 (initiated Jul 17, 2012)
StatusTerminated
ReasonSiemens has received complaints of customers experiencing slopes >1.05 when calibrating Flex(R) reagent cartridge lot 12095AD. This results in a slope failure and prevents the VANC assay from being run on the Dimension Vista(R) System.
Root causeNonconforming Material/Component
ActionThe firm initiated their recall of this product on July 17, 2012 by sending an Urgent Field Safety Notification via FedEx next day delivery to all consignees. The Notice informed consignees of the high calibration slopes with the affected device. Affected consignees were requested to immediately discontinue use and discard any remaining Dimension Vista(R) VANC Flex(R) reagent cartridges, lot 12095AD. Affected customers were provided with replacement cartridges. Customers were to also complete the attached form and fax it back to the number provided. The notification letter should be forwa
Quantity1,894 cartons
DistributionWorldwide Distribution -- US, including Washington, DC and the states of AL, AZ, CA, CO, CT, DE, FL, GA, IA, ID, IL, IN, KS, LA, MA, MD, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NV, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, WA, and WV, and the countries of Canada, New Zealand, Germany, and Puerto Rico.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K061655DIMENSION VISTA FLEX REAGENT CARTRIDGES
Dade Behring, Inc., Jul 10, 2006