Recall Z-2219-2012 ยท LLZ System, Image Processing, Radiological

Siemens Medical Solutions, USA, Inc

syngo Dynamics Version 9.5

DateAug 17, 2012 (initiated Jul 13, 2012)
StatusTerminated
ReasonIf multiple uncompressed images/clips are sent to the syngo Dynamics server, around the same time, it is possible for an image from a patient to be placed in another patient's study.
Root causeSoftware design
ActionOn 07/13/2012, Siemens issued a "Customer Safety Advisory Notice" to affected consignees according to Update Instruction SY062/121S. The notice informs the customers of the potential issue with the identified affected product and provides details regarding the resolution. A software update to relieve the symptom of the issue will be distributed and installed by a Siemens Service Engineer on all affected systems under Update Instruction SY063/121S. A software update to correct this issue will be distributed and installed by a Siemens Service Engineer on all affected systems under Update Ins
Quantity30
DistributionNationwide Distribution -- MA, GA, MO, IA, CA, PA, OR, CO, NY, NJ, WA, UT, OH, WI, MI, KY, and NC.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K102150SYNGO DYNAMICS, VERSION 9.0
Siemens Medical Solutions USA, Inc., Oct 5, 2010