Recall Z-2156-2012 ยท KPR System, X-Ray, Stationary

Philips Healthcare Inc.

Philips MultiDiagnost Eleva with Flat Detector (MD Eleva FD)Model: 708034 Product Usage: Philips X-ray System

DateAug 8, 2012 (initiated Jul 11, 2012)
StatusTerminated
ReasonWhen doing a 3 Dimensional Rotational Angiography (3DRA) the 3D reconstruction does not always start, even when all images have been acquired.
Root causeSoftware design
ActionPhilips issued an Urgent Field Safety Notice dated July 2, 2012 to all affected customers. The letter identified the affected product, problem description, actions to be taken by the customer/user in order to prevent risks for patients or users and actions planned by Philips to correct the problem. For question or support concerning this issue contact your local Philips representative.
Quantity41 units
DistributionWorldwide Distribution - US (nationwide) including the states of CA, CT, MI, MN, MN, NC, NY, NY, VA and the countries of Belgium, Columbia, Czech Republic and Denmark. France Germany Italy Latvia Luxembourg Netherlands Russian Fed. Spain Sweden Switzerland Thailand United Kingdom