ClearanceBrief
Recall Z-2129-2012 ยท DQO Catheter, Intravascular, Diagnostic

ICU Medical, Inc.

Cardiac Catheterization Waste Management System w/1000mL Waste Bag, 2 Fluid Admin Sets Product Usage: For waste managment.

DateAug 2, 2012 (initiated Apr 17, 2012)
StatusTerminated
ReasonThe firm initiated the recall due to a molding defect in a high pressure stopcock where a small percentage of the side ports could become totally occluded.
Root causeNonconforming Material/Component
ActionICU Medical sent an Urgent Medical Device Recall letter dated April 2012 to all affected customers. The letter identified the recalled product, problem and actions to be taken. Customers were instructed to inspect inventory, destroy affected products according to hospital protocol, complete and return verification form and contact ICU Customer Service to order replacement products. The letter states that ICU Medical will credit customer for any product destroyed once the completed verification form is received. This recall should be carried out to the user level and passed on to all those w
DistributionNationwide distribution
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K052865HOSPIRA CATH LAB CONVENIENCE KITS
Icu Medical, Inc., Dec 22, 2005