Recall Z-2132-2012 ยท FKX System, Peritoneal, Automatic Delivery

Baxter Healthcare Renal Div

Extraneal and/or DIANEAL LOW CALCIUM Peritoneal Dialysis Solution in Plastic Container, PL-146 used with Automated Peritoneal Dialysis System/Automated PD Set with Cassette; Manufactured for Baxter Healthcare Corporation, Deerfield, IL 60015 U.S.A. The HomeChoice dialysis system is a personal cycler automated peritoneal dialysis system for pediatric and adult renal patients with fill volumes rangi

DateAug 2, 2012 (initiated Jul 6, 2012)
StatusTerminated
ReasonBaxter discovered that a Baxter technical service representative was providing incorrect instructions for inspection of solution bags for a blocked frangible prior to connection to the Automated PD Set with Cassette.
Root causeEmployee error
ActionBaxter Healthcare Corp. sent "Important Product Information" letters dated July 6, 2012 to the affected home patients and their clinicians, via first class mail on the same date. The letter identified the product, problem and actions to be taken. The customers were instructed to refer to the Extraneal and/or Dineal package insert, Directions For Use - Preparation for Administration and the HomeChoice APD Systems At-Home Guide for proper procedures to connect to a solution line. Customers are instructed to retain a copy of the letter. The customers were also requested to complete the attach
QuantityN/A
DistributionNationwide Distribution includingArizona, California, Ohio, Pennsylvania and Virginia.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K923065BAXTER PERSONAL CYCLER AUTOMMATED PERITONEAL DIAL
Baxter Healthcare Corp, Mar 4, 1994