Recall Z-2121-2012 ยท JLW Radioimmunoassay, Thyroid-Stimulating Hormone

Perkinelmer

Perkin Elmer Genetic Screening Processor (GSP) Product Code: 2021-0010 Product Usage: The GSP instrument is a fully automated, high throughput batch analyzer for time- resolved and prompt fluorescence analysis of samples in microtitration plates. It is intended for in vitro quantitative / qualitative determination of analytes in body fluids. The reagent kits running on the GSP instrument are use

DateAug 1, 2012 (initiated Jun 27, 2012)
StatusTerminated
ReasonContamination from GSP Shaker unit results in decreased signal activity of assay measurements for Genetic Screening for Neonatal IRT, TSH and IRT
Root causeComponent design/selection
ActionPerkinElmer sent an Urgent: Medical Device Recall letter dated June 27, 2012 to all affected customers. The letter identified the affected product, problem and actions to be taken. The letter states that replacement units of the recalled product will be installed as soon as they are available. Customers were instructed to complete and return the response form attached. For further information contact your local PerkinElmer representative or GS. Technical.Support@PerkinElmer.com
Quantity78 units
DistributionWorldwide Distribution - US (nationwide) and the countries of Australia, Belgium, Brazil, Egypt, France, Germany, Netherlands, Norway, Saudi Arabia,Switzerland,
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K090846GSP INSTRUMENT AND GSP NEONATAL HTSH KIT
Perkinelmer, Inc., Sep 3, 2009