ClearanceBrief
Recall Z-2035-2012 ยท JAK System, X-Ray, Tomography, Computed

Philips Medical Systems (Cleveland) Inc

Philips Brilliance 6, 16, 40, 64, Big Bore, iCT and iCT SP are Computed Tomography X-Ray Systems. The devices are Computed Tomography Systems that are intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data taken at different angles and planes

DateJul 19, 2012 (initiated Jul 2, 2012)
StatusTerminated
ReasonPhilips Healthcare was notified that several of the sixteen bolts securing the brackets which support the carbon table top on various models of their Brilliant CT scanner have been found to be defective and the bolt heads may break off.
Root causeProcess control
ActionPhilips sent a Urgent Medical Device Correction letter dated June 28, 2012, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Philips Key Market representatives will distribute the letter internationally. Additionally, a Field Service Engineer will visit each site to inspect and correct the issue, and Philips will be tracking the progressof the correction. For further information, Customers were asked to contact their local Philips representative or Philips Healthcare Office. North American and Canadian consignees sh
Quantity524
DistributionWorldwide Distribution -- USA (nationwide) including the states of AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, IL, IN, KY, LA, MA, MD, ME, MI, MN, MO, MS, NC, MJ, NM, NV, NY, OH, OK, OR, PA, PR, TN, TX, VA, VT, and WI. and the countries of Canada, Argentina, Australia, Austria, Azerbaijan, Brazil,
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K012009MX8000 IDT CT SYSTEM
Marconi Medical Systems, Inc., Sep 13, 2001
K033326PHILIPS PLUS COMPUTED TOMOGRAPHY SYSTEM
Philips Medical Systems, Inc., Oct 29, 2003
K033357ACQSIM-MULTISLICE-CT, MODEL 453567042271
Philips Medical Systems (Cleveland), Inc., Jan 9, 2004
K060937BRILLIANCE VOLUME
Philips Medical Systems (Cleveland), Inc., Jun 5, 2006