ClearanceBrief
Recall Z-2127-2012 ยท JAA System, X-Ray, Fluoroscopic, Image-Intensified

GE OEC Medical Systems, Inc

UroView 2800; MDL number A349855. Product Usage: Designed to provide fluoroscopic and spot-film imaging of the patient during diagnostic, surgical, and interventional procedures. Clinical application may include but are not limited to urologic and endoscopic procedures. The system may be ued for other imaging applications at the physician's discretion.

DateAug 1, 2012 (initiated Mar 27, 2007)
StatusTerminated
ReasonPlease be aware that this is not a new recall. The firm has taken action; but, due to administrative issues this recall is now being classified by the Agency. GE OEC had discovered an issue with the hard disk drives that were installed on OEC device models: UroView 2800; 9800 Plus Digital; MiniView 6800; and FlexiView 8800 Mobile C-Arm. The devices were manufactured between October 15, 2004 and
Root causeSoftware design
ActionPlease be aware that this is not a new recall. The firm has taken action; but, due to administrative issues this recall is now being classified by the Agency. GE OEC sent an Urgent Safety Advisory Notices dated October 11, 2006 and February 21, 2007 via certified mail delivery letter to affected customers. The notices identified the affected products, problem and actions to be taken. Customers were informed that a GEHC OEC field service engineer would be contacting their facility to arrange to have a replacement hard disk drive installed on the affected system(s). Customers were instructe
QuantityPart of a total of 82 units.
DistributionWorldwide Distribution - US (nationwide) and the countries of: Canada, France, Japan, and China.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K993687UROVIEW 2800
Ge Dec Medical Systems, Dec 21, 1999