Recall Z-1972-2012 ยท LGZ Warmer, Thermal, Infusion Fluid

Vital Signs Devices, a GE Healthcare Company

enFlow IV Fluid Warmer (510k name: EFLOW IV FLUID WARMER, MODEL 100, EFLOW DISPOSABLE CARTRIDGE, MODEL 200 AND EFLOW AC POWER SUPPLY, MODEL 120). The Enginivity enFlow TM IV Fluid Warmer is intended for warming blood, blood products and intravenous solutions prior to administration. It is intended to be used by healthcare professionals in hospital, clinical and field environments to help prevent

DateJul 9, 2012 (initiated Feb 20, 2012)
StatusTerminated
ReasonThe EnFlow Warmer Strap does not meet the requirements of the biocompatibility standard (ISO 10993) for products that may have contact with skin for less than 24 hours.
Root causeDevice Design
ActionGE Healthcare sent an "URGENT MEDICAL DEVICE CORRECTION" letter to all affected customers. The letter identifies the product, problem, and actions to be taken by the customers. The letter included customer instructions and a Customer Response Form to be completed and returned to the firm. Contact Customer Service at 1-800-932-0760 for questions regarding this recall.
Quantity11803 cases
DistributionNationwide Distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K060537EFLOW IV FLUID WARMER, MODEL 100, EFLOW DISPOSABLE CARTRIDGE, MODEL 200 AND EFLOW AC POWER SUPPLY, MODEL 120
Enginivity, LLC, Jun 8, 2006