Recall Z-2007-2012 ยท MUJ System, Planning, Radiation Therapy Treatment

Elekta, Inc.

Monaco Product Usage: Planning of radiation therapy

DateJul 12, 2012 (initiated Jun 21, 2012)
StatusTerminated
ReasonIncorrect patient shift directions when the Setup reference dialog is printed out when the DICOM coordinates option is enabled for reports.
Root causeSoftware design
ActionElekta sent an Important Safety Notice to affected customers. The notice identified the affected product, problem, clinical impact and actions to be taken. The notice inform customers of how the issue will be resolved and instructions to follow when it is available.
Quantity38
DistributionUS Nationwide Distribution - including the states of: CA, CT, FL, IL, IN, KY, MI, MN, MO, NJ, NY, NC, OH, OK, OR, PA, WA, and WI
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K110730MONACO RTP SYSTEM
Computerized Medical Systems, Inc., Jun 24, 2011