Recall Z-0564-2013 ยท FMF Syringe, Piston
Becton Dickinson & Company
Case Carton: BD Oral Dispensing Syringe 1 mL Clear with Tip Cap BD, Franklin Lakes, NJ 07417 USA Made in USA www.bd.com BD, Laagstraat 57, B-9140 Temse, Belgium On 100 unit plastic bag: BD oral syringes are intended to dispense oral medications.
| Date | Dec 19, 2012 (initiated Oct 2, 2012) |
| Status | Terminated |
| Reason | BD received a complaint report for mixed syringe tips (oral and non-oral syringe tips) within the same lot. |
| Root cause | Mixed-up of materials/components |
| Action | Becton Dickinson sent a Urgent Product Recall letter/Customer Packing list dated October 2, 2012, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Customers were instructed to review inventory and determine if they have any of the affected product. If so, it should be removed from inventory. The enclosed customer list should be completed as well. Any questions please call 1-201-847-4267. |
| Quantity | 2,786,000 units |
| Distribution | (USA) nationwide |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K980987 | BECTON DICKINSON SYRINGE Bd Becton Dickinson Vacutainer Systems Preanalytic, Jun 11, 1998 |