Recall Z-2348-2012 ยท DTQ Console, Heart-Lung Machine, Cardiopulmonary Bypass
Terumo Cardiovascular Systems Corporation
Safety monitor included with System 8000 Base The Sarns" Modular Perfusion System 8000 is indicated for use for up to 6 hours on the extracorporeal circulation of blood for arterial perfusion, regional perfusion, and cardiopulmonary bypass procedures, when used by a qualified medical professional who is experienced in the operation of this or similar equipment. The safety monitor is indicated to
| Date | Sep 11, 2012 (initiated Aug 12, 2009) |
| Status | Terminated |
| Reason | Terumo Cardiovascular Systems (Terumo CVS) received complaints from our Affiliate, Terumo Dubai that the safety monitors were tripping off. Terumo CVS' investigation determined that 8 monitors were manufactured with an underrated circuit breaker due to an error where an associate pulled an underrated circuit breaker from an incorrect inventory location. Since this time, inventory locations hav |
| Root cause | Mixed-up of materials/components |
| Action | TERUMO Cardiovascular Systems communicated with their affiliate, Terumo Dubai, in 2009 at the time the complaints were reported. No further communication was warranted. For questions regarding this recall call 734-741-6173. |
| Quantity | 8 |
| Distribution | Worldwide distribution: International only-country of: Terumo Dubai, UAE |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K915183 | SARNS 8000 MODULAR PERFUSION SYSTEM 3M Health Care, Sarns, Jul 8, 1992 |