Recall Z-1976-2012 ยท CHL Electrode Measurement, Blood-Gases (Pco2, Po2) And Blood Ph

Abbott Point Of Care Inc.

i-STAT1 Abaxis Analyzer; Abbott Point of Care, Princeton, NJ 08540

DateJul 9, 2012 (initiated Sep 8, 2011)
StatusTerminated
ReasonSome i-Stat Analyzers become uncomfortably hot to touch in the are of the battery. This is a recall extension.
Root causeDevice Design
ActionAbbott Point Care sent an Urgent Product Recall letter dated September 2011 to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to confirm the number of affected product at their facility that were using the Old Battery Carrier. Fill out the affected Buesiness Reply Card and submit their facility information as per the instructions on the card. Upon the receipt of the New Battery Carriers, replace the Old Battery Center with the New Battery Center. Customers were instructed to immediately dispose
Quantity6000-0000-650 units
DistributionWorldwide Distribution - USA (nationwide)
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K001387I-STAT PORTABLE CLINICAL ANALYZER, MODEL 300
I-Stat Corp., Oct 5, 2000