ClearanceBrief
Recall Z-0260-2013 ยท IZW Collimator, Automatic, Radiographic

Shimadzu Medical Systems

R-30H , Diagnostic X-Ray Beam-Limiting Device. In combination with any of below X-ray tubes 0.6/1.2P364DK-85, 0.6/1.2P324DK-125, 0.6/1.2P324DK-85, 0.6/1.2P38DE-85 Device Model # R-30H. Subsequent Product Code: IZX Product Usage: The device is the beam-limiting device for X-ray Radiography.

DateNov 8, 2012 (initiated Oct 28, 2011)
StatusTerminated
ReasonShimadzu is recalling Diagnostic X-ray Beam-Limiting Device, Mobile x-ray system and Diagnostic x-ray tube assembly because they do not comply with the Product Performance Standards, 21CFR 1020.30 (m) Beam quality --(1) Half-value layer (HVL), which was revised in 2006. Certain devices might not comply with the requirements for them in some x-ray generating condition after the revision in 2006.
Root causeNonconforming Material/Component
ActionShimadzu sent an Urgent Voluntary Medical Device Recall - Field Safety Notice letters dated September 19, 2012 to all the affected customers. The letter identifed the description of non-compliance, how to identify the affected products, corrective action plan by Shimadzu and actions to be taken customers. Customers were instructed to contact their local Shimadzu representative or contact Shimadzu National Technical Support at Shimadzu Medical Systems National Technical Support (800) 228-1429.
Quantity665 units
DistributionUS Nationwide Distribution
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K090578SHIMADZU COLLIMATOR R-30H
Shimadzu Corp., Nov 6, 2009