Recall Z-1984-2012 ยท JXG Shunt, Central Nervous System And Components

Integra LifeSciences Corp

Pudenz Flushing Valve with Integral Connectors, catalog number NL850-1330. Integra Neurosciences, Plainsboro, NJ. Device is used for the treatment of hydrocephalic patients when shunting cerebrospinal fluid from the ventricles of the brain.

DateJul 10, 2012 (initiated Apr 4, 2012)
StatusTerminated
ReasonThe pouch label of catalog NL850-1330 indicates an incorrect model and closing pressure of the device. The incorrect label reads: "Model: 12 mm-Medium Pressure" instead of "Model: 16 mm-Low Pressure".
Root causeLabeling Change Control
ActionThe firm initiated a voluntary recall by notifying the two consignees initially via email and then with an "URGENT: MEDICAL DEVICE RECALL" dated April 5th or 9th, 2012. The letter informed them of the labeling error and provided instructions regarding affected product. The International Customer Service number was provided 609-936-5400.
Quantity39 units
DistributionInternational Only Distribution, including the countries of Thailand and Japan.