Recall Z-2088-2012 ยท MKQ Processor, Cervical Cytology Slide, Automated
Tripath Imaging, Inc.
PrepStain PM Kit, T-AG (Catalog 490588); PrepStain PM Kit Europe (Catalog 490671); PrepStain Prep Only PM Kit, T-AG (Catalog 490954); PrepStain PM Kit XLP, T-US-II (Catalog 490424); PrepStain Prep PM Kit XLP T-US-II (Catalog 491118); PrepStain Prep Only PM Kit XLP, T-US-II (490423); PrepStain Prep Only PM Kit, T-US (Catalog 490429); PrepStain PM Kit, T-US (Catalog 490626); PrepStain PM Kit Europe,
| Date | Jul 26, 2012 (initiated Dec 8, 2011) |
| Status | Terminated |
| Reason | Some Prep Stain Kit, 1.2 DiTi Cones have been found to be leaking and the PrepStain Preparation kits contain incorrect preventive maintenance (PM) instructions. |
| Root cause | Nonconforming Material/Component |
| Action | BD Diagnostics sent a letter dated March 2012 to all affected customers. The letter identifies the product, problem and actions that will be taken by the firm. Contact BD Diagnostics Technical Support at 1-877-822-7771, option 2 for questions regarding this letter. |
| Quantity | 1166 units |
| Distribution | Worldwide Distribution-USA (nationwide) including Puerto Rico and the states AL, AK, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, ME, MD, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, WA, WI, WY, and WV and the countries of Belgium, |
| FDA record | Open on fda.gov |