ClearanceBrief
Recall Z-1947-2012 ยท FSA Lift, Patient, Non-Ac-Powered

Arjo, Inc. dba ArjoHuntleigh

Encore Mobile Patient Lifter; manufactured by Arjo Ltd., St. Catherine Street, Gloucester, GL1 2SL, England, U.K.; Models KKB5022-21US and KKB5042-21US. The Encore is designed to be used as a standing and raising aid, it is useful for short transfers from one sitting position to another, and to elevate a patient for toileting, repositioning, changing of incontinence pads or wound dressings, stand

DateJul 6, 2012 (initiated Jun 29, 2012)
StatusTerminated
ReasonArjoHuntleigh has received a limited number of customer reports concerning injuries related to an exposed metal support bracket when the proactive knee pad assembly is removed.
Root causeLabeling False and Misleading
ActionArjoHuntleigh sent an "URGENT FIELD CORRECTION RECALL" notice dated June 25, 2012, to all affected customers via UPS Ground on June 29, 2012. The notice informed the accounts of the potential for severe patient injury if the proactive knee pad support is removed from the lift without the removal of the knee pad support bracket. The accounts were reminded that the entire knee pad and support bracket assembly must be removed from the lift when required to do so for exercise. An additional warning statement with diagrams was included in the notice. The accounts were requested to disseminate the
Quantity2,552 units in the U.S., 4,970 units in the rest of the world
DistributionWorldwide Distribution-USA (nationwide) and the countries of Australia, Austria, Belgium, Bermuda, Canada, Cyprus, Czech Republic, Denmark, Dubai, Estonia, Finland, France, Germany, Gibraltar, Great Britain, Greece, Hong Kong, Iceland, Ireland, Israel, Italy, Japan, Kenya, Korea, Luxembourg, the Ne
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.