Recall Z-1902-2012 ยท LDP Colorimetry, Acetaminophen

Sekisui Diagnostics Llc

MULTIGENT Acetaminophen for use with Architect/Aeroset Cat. No. 2K99-20. Assay is intended for the quantitative determination of acetaminophen in human serum or plasma.

DateJun 27, 2012 (initiated May 29, 2012)
StatusTerminated
ReasonDiscoloration of the acetaminophen enzyme reagent (R1), generates a high calibration factor and erratic control recovery.
Root causeStorage
ActionSekisui Diagnostics Llc sent an "IMPORTANT PRODUCT FIELD CORRECTION" letter dated May 29, 2012 to all affected customers. The letter identifies the product, problem, and actions to be taken by the customers. The letter instructs customers to immediately stop using the affected product and to destroy any remaining inventory. A Confirmation of Notification Form was attached for customers to complete and return via fax to 902-628-6504. Contact the firm at 902-628-0984 for questions regarding this notice.
Quantity1600 kits
DistributionIllinois
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K042330ACETAMINOPHEN-SL ASSAY
Diagnostic Chemicals , Ltd., Jan 7, 2005