ClearanceBrief
Recall Z-1994-2012 ยท GEI Electrosurgical, Cutting & Coagulation & Accessories

PSC Industries Inc

Bayonet Forceps 7.75" (196.9 mm) 0.5 mm Tips Non-Stick Electrosurgical/Bipolar Forceps/Reusable, Model #90-7004 (k) #K982705. The surgical device is package non-sterile into small plastic tubes. Olsen Medical Biopolar Reusable Bayonet Forceps and cords are intended for use in coagulating tissue during surgical procedures.

DateJul 12, 2012 (initiated Jun 1, 2012)
StatusTerminated
ReasonFirm officials reported to CIN-DO that their Bipolar Bayonet Forceps 7.75" are misbranded in that the manufacturer symbol was replaced by the expiry symbol on the packaging.
Root causeLabeling mix-ups
ActionThe firm, Olsen Medical, sent an "URGENT: MEDICAL DEVICE RECALL" letter dated June 5, 2012 to their customers via Certified mail along with representative labeling to illustrate the specific labeling issue associated with the various types and/or sizes of Bayonet Forceps surgical instruments that the customers received. The letter described the product, problem and actions to be taken. The letter instructed the customers to immediately examine their inventory for the presence of the suspect product and quarantine the product, if found; to discontinue use immediately and promptly call (800) 251
Quantity7 units
DistributionWorldwide distribution: USA (nationwide) including states of: CA, CT, FL, GA, MD, MI, NC, OR, SC, and TX and countries of: Canada and Philippines.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K885199#500 SERIES FORCEPS W/CUT/COAGULATE SWITCHES
Olsen Electrosurgical, Inc., Jan 13, 1989