Recall Z-1786-2012 ยท MAX Intervertebral Fusion Device With Bone Graft, Lumbar
Captiva Spine, Inc
Accessories for the Captiva Spine FuseLOX Lumbar Cage. The FuseLOX Lumbar Inserter product is a reuseable component of the Captiva Spine FuseLOX Lumbar product. It is not generally sold as an individually packaged product. Inserter products are shipped as a component of the FuseLOX Lumbar Caddy which is placed in the FuseLox Lumbar Sterilization tray. The sterilization tray is distrib
| Date | Jun 7, 2012 (initiated Apr 27, 2012) |
| Status | Terminated |
| Reason | Captiva Spine, Inc. initiated a recall of FuseLOX Lumbar Implants, Trials, and Inserters after FDA advised the firm that there is inadequate support for implemented design changes to the cleared device Pivotec Lumbar Interbody Fusion Device. |
| Root cause | Finished device change control |
| Action | Captiva Spine sent a Voluntary Correction Notification letter dated May 4, 2012, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to locate and return affected devices. Customers were also instructed to complete and return a Tracking/Verification Form. For questions customers were instructed to call 877-772-5571. For questions regarding this recall call 561-277-9480. |
| Quantity | 8 units |
| Distribution | USA (nationwide) including FL,TX, UT, and NJ. |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
This recall record lists no 510(k) number.