Recall Z-1696-2012 ยท FMF Syringe, Piston

Covidien LLC

Monoject Insulin Safety Syringe 1/2mL 29 X 1/2" Product ID: 8881511136 Indications for Use: Injection of U-1 00 insulin

DateMay 31, 2012 (initiated May 21, 2012)
StatusTerminated
ReasonSafety shield may separate from the syringe, exposing the needle, which could potentially result in a needle stick
Root causeEquipment maintenance
ActionCovidien sent an Urgent Product Recall letter dated May 21, 2012, via Federal Express. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to immediately stop using any affected product. Distributors were requested to notify their customers. Customers were instructed to complete the Monoject Insulin Safety Syringe recalled product return form even if they have no product to return. The form should be faxed to 1-800-895-6140 or e-mailed to sdfeedback@covidien.com. Customer Service questions regarding the recall please use o
Quantity26,655,000 syringes
DistributionWorldwide Distribution - USA (nationwide) and the countries of Canada, Australia, Belgium, and Panama.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K922522MONOJECT SAFETY SYRINGE, 1 CC, 3CC AND 12 CC SIZES
Sherwood Medical Co., Nov 25, 1992