Recall Z-1667-2012 ยท FPK Tubing, Fluid Delivery
MOOG Medical Devices Group
Non-DEHP Epidural Yellow Striped Tubing Admin Set with Vented Vial Spike and 0.2 Micron Filter Packaged Assembly, product code: 340-4144, Method of Sterilization using Ethylene Oxide. The Curlin Ambulatory Infusion Pump System with disposable PVC Administration Tubing Sets and user accessories, provide a means for the volumetric delivery of fluids used in Parenteral, Enteral, Epidural, Subcutaneo
| Date | May 30, 2012 (initiated May 18, 2012) |
| Status | Terminated |
| Reason | Moog Medical Device Group became aware on March 23, 2012 via customer complaint, of a reversed pump segment on an Administration Set. This malfunction could reverse the flow of fluid or medication backwards from what was intended. |
| Root cause | Process design |
| Action | The firm, Moog Medical Device Group, sent an "Urgent Field Safety Notice" dated May 21, 2012 via certified letter and a press release direct to consignees/customers. The notice describes the product, problem and actions to be taken. The customers were instructed to do the following: If purchased by distributor -remove suspect product from inventory; contact your distributor to arrange replacement product and request a shipping return label as per the distributor's process and ship product to their location. If purchased by Moog-remove suspect product from inventory, contact Moog Customer S |
| Quantity | 7,580 sets. |
| Distribution | Nationwide distribution: USA including states of: AZ, CA, FL, IL, KS, MA, MI, MD, NC, ND, NY, OK, OH, and TX. |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K981816 | CUE AMBULATORY VOLUMETRIC INFUSION PUMP Becton Dickinson Curlin, LLC, Dec 4, 1998 |