Recall Z-1538-2012 ยท IZL System, X-Ray, Mobile

Philips Healthcare Inc.

Philips Healthcare Practix Convenio Mobile X-ray system Worldwide Distribution -- Nationwide

DateJun 13, 2012 (initiated Oct 19, 2011)
StatusTerminated
ReasonWhen the system is switched off or when the tube arm is parked in the down position (X-ray generator switch off), a radiation pulse (50 kV, 2.5 pGy) is generated inadvertently. .When the system is switched off or when the tube arm is parked in the down position (X-ray generator switch off), a radiation pulse (50 kV, 2.5 pGy) is generated inadvertently. The failure happened when component supplie
Root causeRadiation Control for Health and Safety Act
ActionPhilips Healthcare sent a proprosed corrective action letter dated May 3, 2012, to FDA. The proposed corrective action plan is to bring the product into compliance with regulatory requirements by issuing a field corrective action which should be completed by Philips by August 1, 2012. For questions regarding this recall call 978-687-1501.