Recall Z-1651-2012 ยท IYE Accelerator, Linear, Medical

Elekta, Inc.

MLCi/MCLi2. Radiation therapy.

DateMay 23, 2012 (initiated Apr 19, 2012)
StatusTerminated
ReasonThe Y2 diaphragm block was found loose.
Root causeNonconforming Material/Component
ActionElekta sent an "IMPORTANT FIELD SAFETY NOTICE" dated April 2012 to all affected customers. The letter identifies the product, problem, and actions to be taken by the customers. The letter instructs customers to carry out the recommended weekly physics check of the X-ray and light field size calibration. If an error is found, customers are instructed to contact their local Elekta service engineer.
Quantity386
DistributionNationwide Distribution including Puerto Rico.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K082122MLCI2
Elekta , Ltd., Aug 29, 2008