Recall Z-1953-2012 ยท IYO System, Imaging, Pulsed Echo, Ultrasonic
Civco Medical Instruments Inc
CIVCO General Purpose Sensor, Non-sterile reusable general purpose electromagnetic sensor for use with 3D Guidance driveBAY, driveBAY 2, trakSTAR, or trakSTAR2 trackers, Ref. 10441627. The firm name on the label is CIVCO, Kalona, IA. The kit contains sensor component part number 410-240. The general purpose sensor within the kit is intended to provide physicians with tools for electromagnetic
| Date | Jul 6, 2012 (initiated May 7, 2012) |
| Status | Terminated |
| Reason | The general purpose electromagnetic sensor was incorrectly calibrated causing potential variability in the projected position of the needle path with the ultrasound device. |
| Root cause | Component design/selection |
| Action | CIVCO sent an Urgent Medical Device Recall letter dated May 3, 2012, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were asked to return all general purpose sensors as the product may not function as intended. To return the affected product customers should contact the OEM Sales Coordinator at 319-248-6521 to schedule the return of their device. Upon receipt of the snesor, CIVCO will arrange for recalibration. Once recalibration is completed, the sensor will be returned to the customer at no charge. Custome |
| Quantity | 30 kits |
| Distribution | Worldwide Distribution - USA including WI, CA, MA, IN, WA, and NJ. Foreign distribution was made to Germany, Canada, Denmark, Hong Kong, Ukraine, Japan, Italy, Korea, and France. |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K092619 | ETRAX NEEDLE SYSTEM, ETRAX VARIABLE ANGLE NEEDLE GUIDANCE SYSTEM, VIRTUTRAX UNIVERSAL TRACKER Civco, Nov 13, 2009 |